Re: [Histonet] Decloaking Chamber (high pressure cooker)
I use a Decloaking Chamber for Chronic Wasting Disease slides, and yes, 20
minutes is the correct setting for that particular test, with a 25 minute
cooldown. Most other IHC tests don't require a lengthy setting on the
Decloakers.
I'd recommend contacting Biocare to do this for you. They came and
installed the settings on my chambers.
Jan Shivers
U of MN Vet Diag Lab
----- Original Message -----
From: "wen eng"
To: "histonet"
Sent: Thursday, October 27, 2005 12:27 PM
Subject: [Histonet] Decloaking Chamber (high pressure cooker)
> Hello,
>
> Does anybody use Decloaking Chamber from Biocare Medical? How should I set
the SP1 temperature and time if I need 20 minutes treatment? The factory
setting is 125 degree 30 seconds and the tech support said 20 minutes is too
long. She suggested me to keep 30 seconds. I still want the suggestion from
here. How do you set your program?
>
> Thanks!